電気筐体圧縮試験装置
ユーザーズマニュアル
適用 LISUN モデル:LS-B21
1. 機器の基本的な理解
1.1 主なパラメータ
1.1.1 Equipment Technical Specifications
| 技術パラメータ | お願い |
| 制御方法 | Touchscreen + PLC Control |
| 圧縮方法 | 電気 |
| Compression Force Value | 50–1000 N, adjustable |
| センサーの精度 | 1%FS |
| Compress Time | Adjustable from 1 to 99 minutes (seconds); once the compression time has elapsed, the system automatically releases pressure and retracts. |
| 変位分解能 | 20 mm |
1.1.2 適用規格
The LS-B21 Electrical Casing Compression Test Device strictly complies with GB/T 17466.1-2019″Electrical Accessories—Installation Boxes and Enclosures for Fixed Electrical Installations for Household and Similar Purposes—Part 1: General Requirements,” specifically Clauses 15.3 and 15.5 (IEC 60670-1:2021 Clauses 15.2 and 15.5), and is used to conduct testing.
It is primarily used in electrical appliance quality inspection laboratories, quality control departments of manufacturers, third-party testing organizations, and for import and export product inspections.
1.2 機器構成の検証
1.2.1開梱
When unpacking the device, handle it gently to avoid scratching the outer casing with sharp tools. After unpacking, first inspect the exterior of the device to ensure there is no visible deformation or dents, and that the buttons are not damaged or loose. If you notice any abnormalities with the device or its accessories, do not turn it on. Please contact us immediately.
1.2.2 本体および付属品
| 数 | お願い |
| ① | LS-B21 Compression Test Device Main Unit |
| ② | 電源コード |
1.2.3 電子文書のダウンロード
We have already sent download links for the electronic user manual, calibration certificate, warranty card, and other documents via email at the time of shipment. Please download them promptly. If you missed the email, you may also contact us directly to request the download links again.
2.安全上の注意
2.1 電気安全
Before connecting the power cord, verify that the supply voltage is AC 220 V ± 10% and the frequency is 50 Hz or 60 Hz;
You must use a three-prong outlet with a protective ground, with a protective ground resistance of ≤4 Ω. Under no circumstances should you cut or disconnect the protective ground wire, as doing so poses a risk of electric shock.
2.2 環境安全
Do not use in environments with flammable or explosive substances, strong electromagnetic interference, high temperatures and humidity, or excessive dust levels;
Operating Environment: Temperature 5°C to 35°C, relative humidity ≤ 65% RH.
2.3 操作手順
Improper operation can easily result in personal injury, equipment damage, and distorted test data; operators must thoroughly read this manual and strictly follow the procedures;
Remove jewelry before performing the procedure, tie back long hair, and do not wear loose-fitting clothing;
During the test, take precautions to prevent test specimen fragments from flying and causing injury;
During the test, do not extend your hands or tools into the fixture’s travel range to prevent crush injuries or accidents;
Operators must not leave their posts without authorization while the equipment is in operation; in the event of an abnormality, immediately shut off the power supply;
Do not perform tests that exceed the load capacity or stroke limit, and do not use damaged or non-standard fixtures.
2.4その他
本マニュアルに記載されていない事項については、慎重に進めるか、弊社までお問い合わせください。
3. 機器の概要
図3-1を参照してください。
| 数 | お願い |
| ① | Power Cord Connection and Power Switch |
| ② | タッチスクリーン |
| ③ | Specimen Mounting Position |

図3-1
4. タッチスクリーンプログラムの概要
4.1 初期画面
図4-1を参照してください。
| 数 | お願い |
| ① | Display current force, displacement, deformation, and maximum force |
| ② | If necessary, enter the sample information. |
| ③ | Test Plan Setup |
| ④ | Display Settings: You can set the units for force values. |
| ⑤ | Curve Display Area |
| ⑥ | Up/Down buttons: Manually adjust the size of the specimen mounting position based on the sample size. |
| ⑦ | Click to clear the current test data |
| ⑧ | Click to return the ram to its initial position |
| ⑨ | Stop Test Button, Start Test Button |

図4-1
4.2 Test Plan Interface
図4-2を参照してください。
| 数 | お願い |
| ① | To test the speed, you can use the default value. |
| ② | Break-induced shutdown: The machine automatically shuts down when the break reaches a set percentage. Optional feature |
| ③ | Test Type: The standard accessories included with this unit are intended for compression testing only. Please select “Compression.” |
| ④ | Start deformation calculation; calculation begins once the actual force reaches a specified percentage of the set force value. Optional setting |
| ⑤ | Shutdown conditions, optional pressure values. Optional settings |
| ⑥ | Start the crush test; once the actual force reaches a certain percentage of the set force, begin calculating the fracture ratio. Optional setting |
| ⑦ | 強制設定 |

図4-2
4.3 Force Value Settings Screen
図4-3を参照してください。
| 数 | お願い |
| ① | Set the required force value |
| ② | Set the pressure duration |
| ③ | Speed reduction settings: When the pressure reaches a set percentage, the speed slows down. Use the default values. |
| ④ | Pressure-activated timing start setting: Timing begins when the actual pressure reaches 100% of the set pressure value. |

図4-3
5. Compression Test Procedure
5.1 Sample Preparation and Placement
Select a specimen of an electrical enclosure that is free of initial cracks, plastic deformation, and in perfect condition. Based on the specimen dimensions and in accordance with Section 4.1, use the up/down keys to adjust the platen to a position slightly higher than the specimen’s height, and place the specimen in the specimen mounting position as described in Chapter 3.
5.2 テストパラメータ設定
Set the compression force and duration in accordance with Section 4.3 and the relevant provisions of the standards.
Clause 15.3 of GB/T 17466.1-2019 (IEC 60670-1:2021 Clause 15.2) requires a force of 500 N applied for 1 minute; GB/T 17466.1-2019, Clause 15.5 (IEC 60670-1:2021, Clause 15.5) requires a force of 50 N/cm² (not exceeding 1000 N) to be applied through 9 mm thick plywood, with a duration of 1 minute.
5.3 Starting the Compression Test
| 数 | お願い |
| ① | Click the “Start Test” button to begin the test |
| ② | Move the press block slowly toward the specimen and apply pressure. |
| ③ | When the actual pressure reaches the set value, the timer starts. |
| ④ | Once the force application time reaches the set duration, the test stops automatically and the load block returns to its original position. |
Note: If the actual pressure does not reach the set value, or if the specimen breaks during the compression process, you may click “Stop Test” to manually terminate the test.
5.4 テスト結果の解釈
After the test is completed, the mounting box and housing must not exhibit any deformation or damage that would affect continued use; if this is the case, the test is considered passed.
6. 機器の日常的なメンテナンスと手入れ
6.1。XNUMX日常のメンテナンス
After each test, clean up any dust or debris from the specimen mounting area to keep the work surface clean and free of debris; wipe down the equipment surface to keep it clean; verify that the clamping block is secure and that the limit stops are not misaligned; apply specialized lubricant to the drive screw to ensure smooth operation.
6.2 Storage Specifications
6.2.1 保管環境の要件
温度:5℃~35℃、相対湿度:65%RH以下、乾燥した、埃のない、腐食性ガスのない環境。
6.2.2 長期保管業務
電源を切り、すべてのケーブルを抜いてください。機器を清掃し、防塵フィルムで包んでください。部品の湿気による損傷を防ぐため、毎月30分間電源を入れてください。
6.3 校正間隔の要件
To ensure the measurement accuracy and operational reliability of the equipment, it is recommended that it be sent periodically to a qualified third-party metrology laboratory for calibration. The recommended calibration interval is 12 months. The specific calibration interval may be determined by the user based on frequency of use, environmental conditions, and quality system requirements.
7.よくある問題のトラブルシューティングと解決
| 障害症状 | 考えられる原因 | 手順 |
| The fixture does not raise or lower | Limit Switch Triggered | Check and adjust the limit switches |
| Force and displacement values are displayed incorrectly | Not reset to zero | Click to reset and try again |
| The machine stops as soon as the specimen is subjected to a load | Parameter Configuration Error | See Section 4.2; do not set unnecessary parameters arbitrarily. |